84.90—Breathing resistance test; inhalation.
(a)
Resistance to inhalation airflow will be measured in the facepiece or mouthpiece while the apparatus is operated by a breathing machine as described in § 84.88.
(b)
The inhalation resistance of open-circuit apparatus shall not exceed 32 mm. (1.25 inch) water-column height (at a flow rate of 120 liters per minute).
(c)
The inhalation resistance of closed-circuit apparatus shall not exceed the difference between exhalation resistance ( § 84.91(e)) and 10 cm. (4 inches) water-column height.