92.109—Analyzer specifications.
(a) General analyzer specifications—
(1) Analyzer response time.
Analyzers for THC, CO2, CO, and NOX must respond to an instantaneous step change at the entrance to the analyzer with a response equal to 95 percent of that step change in 6.0 seconds or less on all ranges used. The step change shall be at least 60 percent of full-scale chart deflection. For NOX analyzers using a water trap, the response time increase due to the water trap and associated plumbing need not be included in the analyzer response time.
(2) Precision.
The precision of the analyzers for THC, CO2, CO, and NOX must be no greater than ±1 percent of full-scale concentration for each range used above 155 ppm (or ppmC), or ±2 percent for each range used below 155 ppm (or ppmC). The precision is defined as 2.5 times the standard deviation(s) of 10 repetitive responses to a given calibration or span gas.
(3) Noise.
The analyzer peak-to-peak response to zero and calibration or span gases over any 10-second period shall not exceed 2 percent of full/scale chart deflection on all ranges used.
(4) Zero drift.
For THC, CO2, CO, and NOX analyzers, the zero-response drift during a 1-hour period shall be less than 2 percent of full-scale chart deflection on the lowest range used. The zero-response is defined as the mean response including noise to a zero-gas during a 30-second time interval.
(5) Span drift.
For THC, CO2, CO, and NOX analyzers, the span drift during a 1-hour period shall be less than 2 percent of full-scale chart deflection on the lowest range used. The analyzer span is defined as the difference between the span-response and the zero-response. The span-response is defined as the mean response including noise to a span gas during a 30-second time interval.
(b) Carbon monoxide and carbon dioxide analyzer specifications.
(1)
Carbon monoxide and carbon dioxide measurements are to be made with nondispersive infrared (NDIR) analyzers.
(3)
The minimum water rejection ratio (maximum CO2 interference) as measured in § 92.120(a) shall be:
(4)
The minimum CO2 rejection ratio (maximum CO2 interference) as measured in § 92.120(b) for CO analyzers shall be 5000:1.
(5) Zero suppression.
Various techniques of zero suppression may be used to increase readability, but only with prior approval by the Administrator.
(6)
Option: if the range of CO concentrations encountered during the different test modes is too broad to allow accurate measurement using a single analyzer, then multiple CO analyzers may be used.
(c) Hydrocarbon analyzer specifications.
(1)
Hydrocarbon measurements are to be made with a heated flame ionization detector (HFID) analyzer. An overflow sampling system is recommended but not required. (An overflow system is one in which excess zero gas or span gas spills out of the probe when zero or span checks of the analyzer are made.
(i) Option.
A non-heated flame ionization detector (FID) that measures hydrocarbon emissions on a dry basis is permitted for petroleum fuels other than diesel and biodiesel; Provided, that equivalency is demonstrated to the Administrator prior to testing. With the exception of temperatures, all specifications contained in Subpart B of this part apply to the optional system.
(ii)
The analyzer shall be fitted with a constant temperature oven housing the detector and sample-handling components. It shall maintain temperature with 3.6 °F (2 °C) of the set point. The detector, oven, and sample-handling components within the oven shall be suitable for continuous operation at temperatures to 395 °F (200 °C).
(v) Premixed burner air.
(A)
For diesel and biodiesel fueled engines, premixing a small amount of air with the HFID fuel prior to combustion within the HFID burner is not recommended as a means of improving oxygen interference (%O2 I). However, this procedure may be used if the engine manufacturer demonstrates on each basic combustion system (i.e., four-cycle direct injection, two-cycle direct injection, four-cycle indirect injection, etc.) that an HFID using this procedure produces comparable results to an HFID not using this procedure. These data must be submitted to the Administrator for his/her approval prior to testing.
(B)
For engines operating on fuels other than diesel or biodiesel, premixing burner air with the HFID fuel is not allowed.
(2) Methane analyzer.
The analytical system for methane consists of a gas chromatograph (GC) combined with a flame ionization detector (FID).
(3) Alcohols and Aldehydes.
The sampling and analysis procedures for alcohols and aldehydes, where applicable, shall be approved by the Administrator prior to the start of testing. Procedures are allowed if they are consistent with the general requirements of 40 CFR part 1065, subpart I, for sampling and analysis of alcohols and aldehydes, and with good engineering practice.
(4)
Other methods of measuring organics that are shown to yield equivalent results can be used upon approval of the Administrator prior to the start of testing.
(d) Oxides of nitrogen analyzer specifications.
(1)
Oxides of nitrogen are to be measured with a chemiluminescence (CL) analyzer.
(ii)
For high vacuum CL analyzers with heated capillary modules, supplying a heated sample to the capillary module is sufficient.