89.307—Dynamometer calibration.
(a)
If necessary, follow the dynamometer manufacturer's instructions for initial start-up and basic operating adjustments.
(2)
Determine the dynamometer calibration moment arm (a distance/weight measurement). Dynamometer manufacturer's data, actual measurement, or the value recorded from the previous calibration used for this subpart may be used.
(3)
When calibrating the engine flywheel torque transducer, any lever arm used to convert a weight or a force through a distance into a torque must be in a horizontal position (±5 degrees).
IT = calibration weight (N) × calibration moment arm (m)
(5)
Attach each calibration weight specified in § 89.306 to the moment arm at the calibration distance determined in paragraph (b)(2) of this section. Record the power measurement equipment response (N−m) to each weight.
(6)
For each calibration weight, compare the torque value measured in paragraph (b)(5) of this section to the calculated torque determined in paragraph (b)(4) of this section.
(7)
The measured torque must be within either 2 percent of point or 1 percent of the engine maximum torque of the calculated torque.
(8)
If the measured torque is not within the above requirements, adjust or repair the system. Repeat steps in paragraphs (b)(1) through (b)(6) of this section with the adjusted or repaired system.
(c)
Optional. A master load-cell or transfer standard may be used to verify the torque measurement system.
(1)
The master load-cell and read out system must be calibrated with weights at each test weight specified in § 89.306. The calibration weights must be traceable to within 0.1 percent of applicable national standards.
(4)
Load the dynamometer to a minimum of 6 equally spaced torque values as indicated by the master load-cell for each in-use range used.
(5)
The in-use torque measurement must be within 2 percent of the torque measured by the master system for each load used.
(6)
If the in-use torque is not within 2 percent of the master torque, adjust or repair the system. Repeat steps in paragraphs (c)(2) through (c)(5) of this section with the adjusted or repaired system.
(d)
Calibrated resistors may not be used for engine flywheel torque transducer calibration, but may be used to span the transducer prior to engine testing.