86.333-79—Dynamometer calibration.
(a)
If necessary, follow the manufacturer's instructions for initial start-up and basic operating adjustments.
(2)
Determine the dynamometer calibration moment arm. Equipment manufacturer's data, actual measurement, or the value recorded from the previous calibration used for this subpart may be used.
Code of Federal Regulations
(4)
Attach each calibration weight specified in § 86.312 to the moment arm at the calibration distance determined in step (2). Record the power measurement equipment response (ft-lb) to each weight.
(5)
For each calibration weight, compare the torque value measured in step (4) to the calculated torque determined in step (3).
(7)
If the measured torque is not within 2 percent of the calculated torque, adjust or repair the system. Repeat steps (1) through (6) with the adjusted or repaired system.
(c) Option.
A master load-cell or transfer standard may be used to verify the in-use torque measurement system.
(1)
The master load-cell and read out system must be calibrated with weights at each test weight specified in § 86.312-79. The calibration weights must be traceable to within 0.1 percent of NBS weights.
(4)
Load the dynamometer to a minimum of 6 equally spaced torque values as indicated by the master load-cell for each in-use range used.
(5)
The in-use torque measurement must be within 2 percent of the torque measured by the master system for each load used.
(6)
If the in-use torque is not within 2 percent of the master torque, adjust or repair the system. Repeat step (2) through step (5) with the adjusted or repaired system.
(d)
The dynamometer calibration must be completed within 2 hours from the completion of the dynamometer warm-up.