86.318-79—Oxides of nitrogen analyzer specifications.
(2)
For high vacuum CL analyzers with heated capillary modules, supplying a heated sample to the capillary module is sufficient.
(b) Option.
The oxides of nitrogen may be measured with an NDIR analyzer system that meets the following specifications:
(1)
The system shall include an NO2 to NO converter, a water trap, and an NDIR analyzer in that order.
(5)
The minimum water rejection ratio (maximum water interference) for the NO NDIR analyzer shall be 5,000:1 (see § 86.321 ).
(6)
The minimum CO2 rejection ratio (maximum CO2 interference) for the NO NDIR analyzer shall be 30,000:1 (see § 86.322 ).